ELTRA‘s CHS-580A is the ideal elemental analyzer for simultaneous determination of carbon, hydrogen and sulfur in organic samples. It features a resistance furnace with vertical ceramic tube and uses ceramic crucibles as sample carriers.
An automatic sample loader allows for analysis of large numbers of samples without operator intervention. The sample loader is available in two versions to accommodate 36 or 130 crucibles. For reliable analysis of acidified samples (TOC analysis), both versions are available in an acid-resistance configuration.
The temperature can be set in steps of 1 °C to a maximum of 1,550 °C. Thanks to sample weights of up to 500 mg and more even inhomogeneous materials can be reliably analyzed. The analyzer can be equipped with up to four independent infrared cells for measuring carbon and sulfur concentrations, allowing for a great variety of applications.
Operation of the CHS-580 A elemental analyzer is simple, fast and convenient. The samples are weighed and registered in the software as manual analysis or analysis via autoloader. Typical sample weights are between 50 and 500 mg. After starting the analysis, the sample is introduced into the furnace and the combustion gases CO2 and SO2 are measured in up to four infrared measuring cells. The results of the elemental analysis can be exported via LIMS, report or text file.
ELTRA CHS-580 A analyzers can be configured to meet a wide range of customer requirements. C, H, S and C/H/S configurations are available with up to 2 infrared cells for each element to cover a wide working range. Robust measuring cells for the reliable determination of sulfur in very high concentrations are also available.
The CHS-580 A series is perfectly equipped for TOC analysis according to DIN EN 15936 with optional acid-resistant autosamplers, as well as the optional TIC module.
An optional autoloader with 130 crucible positions is available for analyses with high sample throughput.
The CHS-580 A elemental analyzer is perfectly suited for measuring carbon, hydrogen and sulfur concentrations in matrices such as cenuşi, materiale de construcţii, coal, cărbune bituminos, ghips, calcar, ulei, materiale de plantare, cauciuc, sand, sol, calamină, tutun, deşeuri, ...
coal
sol
tutun
Elemente măsurate | carbon, hidrogen, sulf |
Probe | organic |
Aliniere cuptor | vertical |
Purtător de probă | creuzete ceramice |
Domeniul de aplicare | agricultură, biologie, chimie / mase plastice, cărbune / centrale electrice, materiale de construcţii, medicină / produse farmaceutice, mediu / reciclare |
Cuptor | cuptor de rezistență (tub ceramic), reglabil pana la 1550º C (trepte de 1º C) |
Metoda de detecție | absorbție în infraroșu |
Numărul de celule IR | 1 - 3 |
Material celula IR | Aur |
Durată tipică de analiză | 60 - 180 s |
Reactivi necesari | hidroxid de sodiu, perclorat de magneziu |
Gaze necesare | oxigen pur 99,5% (2-4 bar / 30-60 psi) aer comprimat (4-6 bar / 60-90 psi) |
Alimentare | 230 V, 50/60 Hz curent max. de încălzire 2000 W |
Dimensiuni (W x H x D) | 55 x 100 x 60 cm 85 x 35 x 45 cm (Autoloader pentru 130 creuzete) |
Greutate | ~ 90 kg (analizor) ~ 15 kg (Autoloader pentru 130 creuzete) |
Echipament necesar | PC, monitor, balanță (rezoluție 0,0001) |
Accesorii necesare | Autoloader (pentru 36 sau 130 de creuzete) |
Accesorii opționale | modul TIC, stabilizator de tensiune 5 KVA |
In the CHS-580A analyzers the sample is burnt in an oxygen atmosphere at temperatures of up to 1,550 °C. The furnace temperature can be freely selected in steps of 1 °C. The combustion gasses (CO2, H2O, SO2) coming from the furnace pass through a dust filter. After the water vapor is chemically absorbed, the dried CO2 and SO2 gas is detected in the additional infrared cells. Depending on the configuration, it is possible to combine up to four infrared cells with different sensitivities.
Supus modificărilor şi erorilor tehnice